Technical developments in multidetector computed tomography (CT) have increased the number of detector rows on the z-axis, and 16-cm wide-area-coverage CT scanners have enabled volumetric scanning of the entire heart. Beyond coronary arterial imaging, such innovations offer several advantages during clinical imaging in the cardiothoracic area. The wide-detector CT scanner markedly reduces the image acquisition time to less than 1 second for coronary CT angiography, thereby decreasing the volume of contrast material and radiation dose required for the examination. It also eliminates stair-step artifacts, allowing robust improvements in myocardial function and perfusion imaging. Additionally, new imaging techniques for the cardiothoracic area, including subtraction imaging and free-breathing scans, have been developed and further improved by using the wide-detector CT scanner. This article investigates the technical developments in wide-detector CT scanners, summarizes their clinical applications in the cardiothoracic area, and provides a review of the recent literature.

Keywords: Multidetector computed tomography, Wide-area detector, Coronary artery disease, Imaging, Technology

INTRODUCTION

Multidetector computed tomography (MDCT) has rapidly developed since the 1980s, and 64-detector CT is an acceptable imaging modality for non-invasive assessment of the coronary arteries. The major difficulties in MDCT for cardiac imaging are the restricted temporal and spatial resolutions and limited z-axis coverage, which could produce stair-step artifacts. In this regard, the development of the wide-area-coverage (16-cm) CT scanner has enabled volumetric imaging of the entire heart within one cardiac cycle, free of stair-step artifacts. These advancements have not only improved the diagnostic performance of coronary CT angiography (CCTA) but also lowered the radiation dose and contrast material volume. Moreover, they offer several advantages over various clinical imaging modalities, such as triple-rule-out (TRO) scan as well as myocardial perfusion and functional imaging. This article summarizes the technical aspects of commercially available wide-detector CT scanners and suggests their potential clinical applications and benefits in cardiothoracic imaging.

Technical Aspects of the Wide-Detector CT Scanner System

Development of Wide-Detector MDCT

Clinical interest in the application of CT for imaging the coronary arteries started in 1998, with the introduction of “4-detector-row” CT scanners. However, these early MDCT models showed limitations in performing coronary angiography; therefore, their use in cardiac applications was confined to coronary calcium scoring, a technique originally established on electron-beam CT scanners (). Since the development of 16-detector-row CT scanners, CCTA became feasible with the retrospective electrocardiogram (ECG)-gating data-acquisition method; nevertheless, its use is still limited to routine clinical practice (). Following the introduction of 64-detector-row CT scanners, the diagnostic accuracy of CCTA improved, and it became a clinically feasible technique owing to the smaller detector element size (0.5–0.65 mm) and faster rotation time (350–420 ms). Although 64-detector-row CT represents a marked improvement over conventional MDCT, especially in cardiac imaging, its craniocaudal coverage without gantry movement is typically only 20–40 mm, which limits the width of coverage on cine imaging in the craniocaudal direction.

To overcome this disadvantage, the development of wide-area detectors that would enable > 40-mm coverage per rotation was led by Canon Medical Systems (formerly, Toshiba Medical Systems), Japan. They developed a prototype 256-detector-row CT scanner in 2005, which was designed to allow wide craniocaudal coverage, imaging approximately 100 mm in one rotation (,,). Thereafter, the first commercially available wide-range detector scanner was released in November 2007 (Aquilion One; Canon Medical Systems, Otawara, Japan). It had a detector configuration of 320-detector rows × 0.5-mm detector element width, which enabled 16-cm coverage and a 350-ms gantry rotation time at a single focal spot. This configuration allowed three-dimensional (3D) volumetric whole-heart imaging during the diastole of one R-R interval; however, the standard temporal resolution was approximately 175 ms (one-half the gantry rotation time), which remains significantly longer than the 33 ms achieved using catheter coronary angiography. This 320-detector-row CT scanner shows promising temporal homogeneity and is free of the stair-step artifacts noted when performing volume scans, but it often suffers from cone-beam geometry-related artifacts and insufficient temporal resolution. Since 2011, Canon Medical Systems has adopted the dual focal spot (two alternating z-focal spot positions) technique and a faster gantry rotation time (275 ms). Therefore, the latest-generation wide-detector scanners allow imaging of 640 slices with a 137-ms temporal resolution. GE Healthcare introduced a new volume CT scanner (Revolution CT; GE Healthcare, Waukesha, WI, USA) in 2015; it had 256 detector rows (512 slices), which enabled 16-cm coverage, with a 0.625-mm detector element width, and a gantry rotation speed of 280 ms. In addition, the integration of new algorithms allowed the reduction of cone-beam artifacts (,).

Philips Healthcare introduced a wide-coverage scanner in early 2008 (Brilliance iCT; Philips Healthcare, Cleveland, OH, USA), which had 128 detector rows that enabled 8-cm coverage with a 0.625-mm individual detector row width and a 270-ms gantry rotation time. The third-generation dual-source CT scanner SOMATOM Force developed by Siemens Healthineers (Forchheim, Germany) has 6-cm coverage (96-detector rows × 0.6-mm detector element width) but enables fast scan coverage because of the use of double helices and higher pitch values owing to the presence of two X-ray tubes and detector arrays. Generally, the length of the cardiac area that has to be covered in a coronary CT scan is typically around 120–140 mm (); therefore, in this review, we will focus on the 16-cm-coverage CT scanners (256–320 detector rows) that enable volume scans of the entire heart within one cardiac cycle (Table 1).

Contrast Media and Radiation Dose

See article.

Effect of Heart Rate and Image Quality

See article

Applications of Wide-Detector CT in Cardiothoracic Imaging

Volumetric Cardiac Imaging for Function and Perfusion

Although CCTA is extensively used to evaluate the coronary arteries, it can also provide 3D volumetric information regarding the cardiac chambers or the myocardium. In retrospective ECG-gated CCTA, volumetric left ventricular (LV) functional parameters can be measured simultaneously with the LV mass and volume. The quantification of LV function using CCTA has been validated in many prior studies through a comparison with cardiac MRI, which is considered the gold standard for assessing LV function (,). The conventional method for assessing LV function with a 64-detector-row CT scanner is a helical scanning technique with retrospective ECG gating, which covers the entire heart in multiple heartbeats; however, this approach is associated with the risk of motion artifacts due to arrhythmias or breathing (). The wide-detector CT scanner allows volumetric imaging for a 16-cm z-axis coverage with one heartbeat and no time delay; thus, it can acquire both isophasic and isochronic CT data and can yield volumetric and functional measurements of the cardiac chamber with greater accuracy and less radiation exposure (,).

Although CT has high sensitivity and negative predictive value for the detection of significant coronary stenosis, the evidence of coronary stenosis is not predictive of inducible ischemia; while the presence and extent of myocardial ischemia are important determinants of adverse outcomes (,,). Groothuis et al. () compared CCTA with MR myocardial perfusion imaging and found that only 42.3% of patients with obstructive coronary arterial disease on CCTA showed myocardial ischemia on MR perfusion. Thus, additional functional assessments, such as fractional flow reserve (FFR) measurements, cardiac single-photon emission computed tomography (SPECT), stress echocardiography, or stress perfusion MR, are needed to determine the indications for revascularization (,,). Recent advances in CT technology have enabled myocardial CT perfusion imaging with high temporal resolution and a reasonable radiation dose. Two approaches in CT perfusion imaging—static and dynamic—are currently available. Static CT perfusion imaging refers to the assessment of myocardial enhancement acquired at a single time point of the first pass of the contrast material bolus. Since Kurata et al. () first reported static myocardial CT perfusion in 2005, the application of CT for assessing myocardial ischemia has become widespread, and several clinical studies, including multicenter/multivendor trials, have demonstrated the feasibility of static CT perfusion imaging (,,). However, the success of static CT perfusion is highly dependent on contrast material bolus timing and cardiac motion, and the single-shot technique cannot provide quantitative values of myocardial perfusion (,). In contrast, dynamic CT perfusion is assessed on the basis of myocardial enhancement at multiple time points of the first pass of the contrast material, which is robust in terms of bolus timing and allows for fully quantitative analysis of myocardial perfusion (). However, the clinical implementation of dynamic CT perfusion assessment has been challenging because of the limited z-axis coverage of 64-detector-row CT scanners and the relatively high radiation dose inherent to repetitive image acquisition. Although coverage of the LV myocardium is possible using narrower detector scanners in the shuttle mode, the sampling rate for a given part of the myocardium is not high (every second to fourth heartbeat, depending on the heart rate) and the myocardium is temporally different in the cranial and caudal parts (,). Wide-detector CT scanners enable dynamic whole-heart volumetric scanning, thereby eliminating the time discrepancy between the base and apex of the heart seen with helical imaging. This opens up the possibility of determining myocardial perfusion precisely (). In addition, using a wide-detector CT system has several other advantages, including temporal uniformity in image acquisition as well as the possibility of targeting a specific portion of the contrast bolus and reducing the radiation dose and contrast material volume (). The “CORE-320” multicenter trial group, which was established to support coronary artery evaluation using 320-row MDCT angiography, published several papers regarding the diagnostic accuracy of 320-row MDCT for detecting myocardial perfusion deficits with perfusion CT scanning and compared it with the reference standard of SPECT myocardial perfusion imaging (,,,).

Transluminal Attenuation Gradient

CCTA is an established noninvasive method for anatomic assessment of coronary stenosis; however, it has limited specificity for densely calcified lesions and provides no information about their hemodynamic significance. Transluminal attenuation gradient (TAG), defined as the linear regression coefficient between luminal attenuation and axial distance, has been continuously developed since 2010 in an attempt to improve the diagnostic performance of CCTA (,). The methodology of TAG is similar in most studies: on the cross-sectional images perpendicular to the centerline for target coronary artery, the mean luminal attenuation is measured at 1–5-mm intervals, from the ostium to the distal level where the lumen area fell below 2 mm2. Then, TAG is determined from the change in HU per 10-mm length of coronary artery (Fig. 1). This simple technique retains the functional significance of coronary stenosis without additional radiation exposure or contrast material.

Fig. 1

A. Coronary CT angiography showed LAD with significant dense calcified plaque burden that was suspicious for severe obstructive lesion (arrows). B. Invasive coronary angiography showed severe luminal narrowing of LAD (arrows). C. TAG of LAD measured with semi-automated method by using dedicated computer software (Canon Medical Systems). Mean luminal radiologic attenuation (HU) was measured at 1-mm intervals, from ostium to distal level (cross-sectional area < 2 mm2). TAG was −24.98, which indicates significantly low value (generally accepted cutoff value: −15 to −18). CT = computed tomography, HU = Hounsfield units, LAD = left anterior descending coronary artery, TAG = transluminal attenuation gradient

The diagnostic performance of TAG was reported to differ among CT scanners with varied longitudinal coverage, and coronary attenuation variability may occur because of the lack of temporal uniformity among 64-detector-row CT scanners (,,,). Chow et al. () attempted to overcome this inherent limitation of 64-detector-row CT scanners; they suggested corrected coronary opacification (CCO), which was calculated as the quotient of the mean HU in the coronary artery and aorta (CCO = coronary artery attenuation/aortic attenuation), to normalize the attenuation values of the coronary lumen to the aorta. Theoretically, whole-heart coverage with single-rotation, single-beat imaging using a wide-detector CT scanner may be ideal for TAG measurement. Several studies have demonstrated that TAG with a 320-detector-row CT scanner has a higher accuracy than does CCTA alone, owing to improvements in specificity and positive predictive value (,).

However, clinical validation studies have reported conflicting results despite performing TAG using wide-detector CT scanners, and Park et al. () suggested that TAG may be affected by changes in the coronary luminal diameter (i.e., the transluminal diameter gradient [TDG]), implying that TAG may be a secondary result of differences in diameters. A recently published study suggested that TAG, TAG-CCO, and TDG do not discriminate between vessels with or without ischemia as defined by quantitative positron emission tomography or FFR ().

Triple-Rule-Out Scan and Pulmonary Vascular Imaging

TRO CT protocols can serve as cost-effective diagnostic tools for simultaneously evaluating the coronary arteries, aorta, and pulmonary arteries in patients with acute chest pain; however, these protocols require larger radiation exposures and contrast material volumes to obtain sufficient opacification of the thoracic vessels and appropriate diagnostic image quality. Because of the use of detailed-pitch retrospective ECG-gated scanning for the coronary arteries and the overlapping coverage of the same anatomic area for ruling out different causes, TRO CT examinations involve higher radiation doses and larger contrast material volumes owing to the longer scan time compared to that of CCTA. In addition, approximately 10% of TRO CT images are non-diagnostic in coronary arterial evaluation because of inadequate image quality, especially when using the conventional 64-detector-row CT scanner (,).

Various CT techniques have been proposed to reduce the radiation dose and contrast material volume for TRO CT, including the lower kVp protocol and prospective ECG-triggered high-pitch dual-spiral technique with dual-source CT. However, these techniques could only be applied in specific patient populations: the low kVp protocol in non-obese patients and the high-pitch mode in patients with a low regular heartbeat (below 65 bpm) (,,).

Wide-detector CT scanners enable whole-chest CT angiography with two to three axial volume scan acquisitions, which reduce stair-step artifacts and lead to a lower radiation exposure (Fig. 2) (,,). Kang et al. () compared the wide-volume and helical modes of TRO CT using a 320-row-detector CT scanner and reported that the use of the wide-volume scan mode reduces the radiation dose by around 60% and yields similar image quality to that of helical scans; moreover, even patients with fast or irregular heart rates can be scanned using the wide-volume scan mode and diagnostic images of the coronary arteries can be obtained. Recently, a similar study by Chen et al. () using a 256-detector-row CT scanner also reported results consistent with those obtained using a 320-detector-row CT scanner.

Fig. 2

A. Detector width is set at 14–16 cm to cover half of entire thorax, and whole chest is covered with two axial volume scan acquisitions. B. Whole-chest CT scan acquired using two axial volume scans that are automatically stitched immediately after reconstruction. C. Data from second scan are separately reconstructed for coronary angiography. Multifocal low-density filling defects in both lobar and segmental pulmonary arteries suggesting pulmonary arterial thromboembolism (arrows in B). No definite abnormality is observed in thoracic aorta and coronary arteries. RCA = right coronary artery

Transcatheter Aortic Valve Implantation Evaluation

Transcatheter aortic valve implantation (TAVI) is a valid alternative therapeutic procedure for patients with symptomatic severe aortic stenosis who are ineligible for surgery or are high-risk surgical candidates (,). Currently, the indications for TAVI have expanded to include patients at intermediate risk for conventional surgical valve replacement (,). While CT was initially used primarily for the assessment of peripheral access, the role of CT has grown substantially, and it is now the gold standard tool for annular sizing, determining the risk of annular injury and coronary occlusion, and providing appropriate projection angles for fluoroscopic procedures (). Using a conventional 64- or a 128-row-detector scanner, the TAVI planning CT protocol usually requires two separate acquisitions: a retrospective ECG-gated helical acquisition of the aortic root, followed by an ungated helical acquisition of the aorta and iliofemoral arteries (,). This separate acquisition approach requires the injection of two separate boluses of the contrast material, resulting in the administration of a considerable amount of iodine. However, most TAVI candidates are elderly patients who frequently have compromised renal function (,,). Moreover, accurate measurement of the aortic root annulus maximum size is critical for TAVI planning; thus, image acquisition with retrospective ECG gating to cover the entire cardiac cycle should be considered (). Therefore, the application of the prospective ultra-high-pitch mode with dual-source scanners seems inappropriate for TAVI planning (,).

Many authors have suggested various acquisition protocols for TAVI to reduce the contrast material volume and radiation dose using the wide-detector CT scanner (,,), and the suggested methods could be roughly divided into two types: 1) axial ECG-gated volume acquisition for the aortic root and heart followed by ungated helical acquisition for the abdomen and pelvis (,) and 2) axial ECG-gated volume acquisition for the aortic root and heart followed by three or four ungated volume acquisitions for the abdomen and pelvis (). According to previous studies, wide-detector CT scanners enable 30–44% reduction in the iodine load and 32–48% reduction in the radiation dose in TAVI planning, while maintaining or providing better image quality than that of the conventional protocol using 64-detector-row CT scanners (,).

Many authors have suggested various acquisition protocols for TAVI to reduce the contrast material volume and radiation dose using the wide-detector CT scanner (,,), and the suggested methods could be roughly divided into two types: 1) axial ECG-gated volume acquisition for the aortic root and heart followed by ungated helical acquisition for the abdomen and pelvis (,) and 2) axial ECG-gated volume acquisition for the aortic root and heart followed by three or four ungated volume acquisitions for the abdomen and pelvis (). According to previous studies, wide-detector CT scanners enable 30–44% reduction in the iodine load and 32–48% reduction in the radiation dose in TAVI planning, while maintaining or providing better image quality than that of the conventional protocol using 64-detector-row CT scanners (,).

Fig. 3

Images are obtained using conventional helical acquisition method with sedation (A) and single-axial volume scan method without sedation (B) using 16-cm wide-detector CT scanner. A. Helical mode images show severe motion artifacts with blurring (arrowheads), especially near diaphragm (arrows). B. Axial volume scan images show relatively good image quality with few respiratory motion artifacts near diaphragm (arrows) and more clearly defined lung parenchymal structures, bronchi (arrowheads), vessels, and lung fissures.

In addition, wide-detector CT scanners have a potential use in free-breathing dynamic airway evaluation in children suspected of having airway abnormalities such as tracheobronchomalacia (). Traditionally, airway abnormalities have been diagnosed using bronchoscopy, which has significant disadvantages: it is invasive, requires general anesthesia, and has limited application in infants or small children (,). Moreover, stenosis may impede the passage of the bronchoscope and endoscopy may not enable evaluation of the airway distal to a severe stenosis. Therefore, attempts have been made to use MDCT as an alternative evaluation tool for airway assessment (,,,,). However, 64-detector-row MDCT shows an inherent limitation in z-axis coverage, and hence, different airway segments are imaged at varying phases of respiration for acquiring images of the entire central airway (). The development of a wide-detector CT scanner (with 16 cm of anatomical coverage) facilitates dynamic imaging of the entire airway at all phases of respiration (,,,). Wide-detector CT scanners also allow for volumetric assessment of the airway during free breathing, without anesthesia and without having to move the table through the gantry. Generally, the protocol for dynamic pulmonary scanning involves four to five gantry rotations within a total time of 1.4 seconds and the reconstruction of 8–10 phases (,,). Since the initial study on dynamic airway volume CT in 2010 (the mean effective radiation doses were 5–8.2 mSv) (), researchers have been trying to reduce the radiation dose using low-voltage, low-current, and iterative reconstructions (,,,), and the mean effective dose reported in the most recent study by Andronikou et al. () is 1.0 mSv.

Challenging Techniques in Cardiothoracic Imaging: Free-Breathing Scans and Subtraction Imaging

CCTA examinations are usually performed in a breath-holding state to reduce the artifacts caused by respiration. However, CCTA cannot be performed successfully in some patients who cannot hold their breath (e.g., patients who are deaf or have hearing weakness, those with severe pulmonary disease, and those who are uncooperative). Sometimes, the heart rate increases and the variations caused by breath-holding may result in poor image quality. Wide-detector CT scanners are capable of performing one cardiac cycle axial volume scans for CCTA with very short acquisition times without table motion; thus, cardiac shifting caused by respiration during the scan should be insignificant. Kang et al. () reported the feasibility of performing CCTA during free breathing by using a first-generation 320-detector-row CT scanner in patients with heart rates lower than 75 bpm; they also reported that axial volume scans for CCTA resulted in noninferior image quality compared to that acquired using breath-holding CCTA. With the improvements in gantry rotation time and the application of motion-correction algorithms, the latest version of wide-detector CT scanners (256 detector rows) enables the acquisition of images with excellent quality and remarkable diagnostic accuracy despite the high heart rate during the scan (,).

Given the improvements in CT technology, CCTA can exclude stenotic coronary arterial lesions with high diagnostic accuracy. Despite these technological advances, which have improved both spatial and temporal resolutions, extensive coronary calcification or stents are still frequently uninterpretable or overestimated in terms of their severity due to blooming artifacts (,). Volumetric CCTA acquisition using a wide-detector CT scanner allows coronary subtraction technology, which is a recently developed method that removes calcium and stents from CCTA images. Briefly, the subtraction images are obtained using corresponding non-contrast and contrast datasets, and both of them are registered and subtracted. The result is a 3D volume in which coronary calcifications/stents have been removed, leaving the contrast-enhanced blood in the lumen as the only high-intensity material (,). Since Yoshioka and Tanaka () reported the initial experiences with subtraction CCTA using a 320-detector-row CT scanner in 2011, many researchers have attempted to develop better acquisition methods or subtraction algorithms. The most important factors for successful achievement of subtraction CCTA should be accurate registration of the two datasets; therefore, most current studies are based on the single breath-hold method to decrease the likelihood of misregistration between the two datasets (,,,,).

The feasibility of subtraction CCTA has been verified by several studies, and subtraction CCTA allows for a significant improvement in diagnostic accuracy owing to its higher specificity and positive predictive value for the assessment of coronary artery stenosis than those of conventional CCTA alone (,,,,,,). However, the proportion of segments excluded because of misregistration is still high even in the most recently published study, which reported misregistration in approximately half (53%) of all target segments (). Manual correction can be considered in case of significant lesions, but it is highly time-consuming. Moreover, although the breath-holding times of this single breath-hold method have been shortened to approximately 15–18 seconds, this breath-holding time remains problematic in some patients (,). In highly selected patients for whom CCTA can be challenging because of an extensive calcium load or inability to tolerate breath-holding, with successful registration, subtraction CCTA may be helpful in the current technique.

CONCLUSION

Wide-detector CT scanners provide advantages in terms of improvement in image quality and reduction of the radiation dose and contrast material volume in routine CCTA examination than in conventional 64- or 128-detector-row scanners. Moreover, given their unique characteristic of z-axis coverage, several new technologies (such as myocardial function/perfusion, TAG, or subtraction imaging) have become more feasible and better reflect the exact physiological status of patients. Further development of wide-detector CT scanners, especially for improvements in temporal and spatial resolutions (with faster rotation times and thinner individual detector widths), may provide great clinical advantages in the future.